Anvisa has launched four public consultations to gather input on a partial review of the regulatory framework for the processing of medical devices, which is expected to update regulations issued nearly two decades ago, such as Anvisa RDC No. 156/2006, Anvisa RE No. 2,605/2006, Anvisa RE No. 2,606/2006, and Anvisa RDC No. 15/2012. The consultations are as follows:
- Anvisa CP No. 1,420/2026: Addresses the processing of medical devices in healthcare facilities, assigning to the healthcare facility the steps under its management and the decision to make the device available for clinical use.
- Anvisa CP No. 1,421/2026: Addresses processing companies, assigning them responsibility for the steps they perform, record-keeping, traceability, and related communications.
- Anvisa CP No. 1,422/2026: Addresses the validation of processing protocols.
- Anvisa CP No. 1,423/2026: Addresses the classification of medical devices as single-use or reusable.
Anvisa CP No. 1,420/2026 – Processing in Healthcare Facilities
Anvisa CP No. 1,420/2026 proposes that medical devices may only be processed in healthcare facilities within a Material and Sterilization Center (CME) classified as Class 1 or Class 2. Under the text, a Class 1 CME consists of an area designated for processing in an isolated examination room or in a Type I endoscopy unit, where processing may be carried out on a partitioned workbench shared with patient care activities.
The proposal requires every healthcare facility with a Class 2 CME to establish and maintain a Processing Committee composed of, at a minimum, the CME’s technical director and representatives from patient care, quality management, patient safety, infection control, clinical engineering, and procurement or supplies, replacing the current threshold of 500 surgeries per month. The draft also requires the CME to (i) establish and maintain a quality management system, (ii) conduct internal audits at scheduled intervals, at least annually, and (iii) subject non-critical medical devices to at least low-level disinfection, whereas currently only cleaning is required. Under the proposal, the healthcare facility must also maintain a documented procedure to identify, record, evaluate, investigate, and report adverse events, technical complaints, and nonconformities potentially related to the processing of medical devices. The draft also regulates the use of mobile processing units.
With regard to outsourcing, Anvisa CP No. 1,420/2026 proposes that the healthcare facility and the processing company enter into a service agreement setting out each party’s responsibilities for all stages of processing, as well as the procedures for communication, investigation, information sharing, and reporting of adverse events, technical complaints, and nonconformities. In addition, the healthcare facility may only engage a processing company that meets the applicable health requirements, including a local health license and an AFE, when applicable, and that performs all stages of processing except pre-cleaning, which remains the responsibility of the healthcare facility. The draft also provides that the healthcare facility is jointly responsible for the safety of the processing carried out by the contracted processing company and must, among other measures, conduct quality audits of that company.
Anvisa CP No. 1,420/2026 further prohibits the CME from processing a medical device classified as single-use, except when the healthcare facility, through a formal and technically justified institutional decision, assumes health responsibility for processing the device after its use on a patient, provided that the following conditions are cumulatively met: (i) approval by the Processing Committee or an equivalent body, (ii) a documented risk assessment, (iii) a specific, validated, and documented processing protocol, (iv) documented evidence of safety, performance, integrity, functionality, cleaning, disinfection, or sterilization, (v) individual traceability, and (vi) a defined maximum number of processing cycles and objective disposal criteria. Under the proposal, the processing of medical devices whose reuse is prohibited by specific regulation remains prohibited under any circumstances.
Anvisa CP No. 1,420/2026 also establishes that using a proprietary processing protocol does not relieve the healthcare facility of responsibility for the processing decision and for releasing the device for clinical use, without prejudice to the responsibility of the manufacturer or marketing authorization holder for the adequacy, accuracy, and timeliness of the information in the labeling and instructions for use.
Anvisa CP No. 1,421/2026 – Processing Companies
Anvisa CP No. 1,421/2026 proposes that processing companies hold a Company Operating Authorization (AFE) issued by Anvisa for the processing of medical devices and a health license issued by the local health surveillance authority.
Under the proposal, processing services must be formalized in a contract between the processing company and the contracting healthcare facility, covering, at a minimum, the responsibilities of each party, the devices or groups of devices covered, the steps performed, requirements for pre-cleaning, transportation, and traceability, the reporting and investigation of adverse events, technical complaints, and nonconformities, information sharing, notification responsibilities, returns, disposal, audits, access to records, and criteria for suspending the services.
The text also provides that the processing company is responsible for the processing steps performed at its facilities, while pre-cleaning must be performed by the healthcare facility before the devices are sent to the processing company. When a device classified as single-use is processed after its use on a patient, the processing company is responsible for the stages it performs, record-keeping, traceability, incident investigation, and prompt reporting of adverse events, technical complaints, and nonconformities to the contracting healthcare facility, without prejudice to the healthcare facility’s responsibility for the decision to use the processed device. The draft also prohibits the processing company, among other actions, from providing processing services to consignors or distributors of medical devices and from selling used medical devices.
Anvisa CP No. 1,421/2026 proposes that the processing company maintain traceability procedures and records for all medical devices, including control of the number of processing cycles, so that each device can be linked to the originating healthcare facility and patient, under mechanisms defined in the contract and in compliance with applicable personal data protection laws. The company must also maintain procedures ensuring that data such as model, reference, lot, serial number, or UDI can be linked for the purposes of the manufacturer’s field actions. The proposal also requires the processing company to establish and maintain a Master Validation Plan or an equivalent document.
Anvisa CP No. 1,422/2026 – Validation of Processing Protocols
Anvisa CP No. 1,422/2026 establishes that healthcare facilities and processing companies responsible for processing medical devices must develop, validate, implement, monitor, and review processing protocols in accordance with the instructions for use provided by the manufacturer or marketing authorization holder, when available and applicable.
Under the proposal, groups of devices with similar attributes may be established, and validation for each group must be based on the “worst-case” criterion. Anvisa CP No. 1,422/2026 further proposes that validated protocols remain under continuous monitoring to ensure their validated status is maintained, and that they be revised or replaced, with validation of a new protocol or revalidation, in situations such as changes to the technologies used, the facilities, or the manufacturer’s instructions for use. For medical devices classified as critical or single-use, the draft requires individual traceability.
Anvisa CP No. 1,423/2026 – Classification of Single-Use and Reusable Devices
The proposal does not apply to in vitro diagnostic medical devices or to software as a medical device. The current model, which distinguishes between products prohibited from processing and those that may be processed based on the list in Anvisa RE No. 2,605/2006, would be replaced by classifying each device as either single-use or reusable.
Anvisa CP No. 1,423/2026 proposes that the marketing authorization holder be responsible for reporting this classification, as determined by the manufacturer based on the device’s design, and keeping it up to date, and that any change in classification be requested by the marketing authorization holder through an amendment with immediate effect. Under the proposal, Anvisa may order a change in classification at any time for reasons of public health risk.
The text prohibits the reuse of a medical device that (i) comes into contact with cytostatic drugs or radiopharmaceuticals, (ii) incorporates drugs into its composition, (iii) is implantable, (iv) is partially or fully absorbable, (v) has a battery that cannot be replaced or that poses a risk of malfunction after processing, or (vi) has been used in a procedure on a patient with confirmed or suspected Creutzfeldt-Jakob disease.
Anvisa CP No. 1,423/2026 also proposes that only new single-use medical devices in the manufacturer’s original packaging be permitted for import, and prohibits the sale and import of used medical devices or devices processed for reuse, as well as the import, sale, distribution, or donation of medical devices whose classification does not match the one stated in the registration or notification filed with Anvisa.
The proposal further provides that sterilization packaging and chemical and biological indicators must comply with the NBR ISO 11607, NBR ISO 11138, and NBR ISO 11140 series of technical standards, as applicable, and are not subject to notification or registration as medical devices. Under the proposal, manufacturers and importers of these products would not need an AFE, and their imports would not be subject to the procedures set forth in Anvisa RDC No. 81/2008.
Transition Rules
The draft regulations do not establish a single transition period, which will require coordinated planning across the supply chain. The proposed timelines are as follows:
| cp | Transition Deadline |
| Anvisa CP No. 1,420/2026 | 24 months from the date of publication |
| Anvisa CP No. 1,421/2026 | |
| Anvisa CP No. 1,422/2026 | |
| Anvisa CP No. 1,423/2026 |
90 days after publication Five years from the effective date for marketing authorization holders to request classification changes and update their dossiers |
Next Steps
The public consultations will be open until January 28, 2027, and are intended to gather input on transition costs, compliance deadlines, AFE requirements for processing companies, and the implementation of validation and traceability requirements.
The Life Sciences and Healthcare practice can provide further information on this topic.